SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Company
Beckman Coulter Inc.
Recall Initiated
November 10, 2021
Recall Number
Z-0399-2022
Quantity
1433
Firm Location
Brea, CA

Reason for Recall

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

Distribution

US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand

Lot / Code Info

Lot: M106450, UDI: 15099590575113

Root Cause

Nonconforming Material/Component

Action Taken

On 11/10/21 recall notices were sent to customers. Customers with questions are encouraged to contact the Customer Support Center: http://www.beckmancoulter.com, 1-800-854-3633 in the United States and Canada. Customers were asked to take the following action: 1) Discontinue use of affected cholesterol reagent and follow the product replacement instructions. 2) Review the Safety Data Sheet (SDS) and discard according to your laboratory s Standard Operating Procedures and/or your local regulations. 3) Retrospective review is left to the discretion of the Laboratory Director. 4) Share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 5) Complete and return the Response Form. For customers in the United States, if you need replacement product, complete the Replacement Order Form.

More recalls by Beckman Coulter Inc.

View all recalls by this company →