RecallDepth

Arietta 850 Ultrasound- Software Version: V1.0.0 through V4.1.3

Company
Fujifilm Healthcare Americas Corporation
Recall Initiated
October 14, 2021
Recall Number
Z-0190-2022
Quantity
38 units
Firm Location
Twinsburg, OH

Reason for Recall

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

Distribution

Nationwide

Lot / Code Info

Serial Numbers: 205W3386 (01)04562122958974(11)171114(21)205W3386; 206F7900 (01)04562122958974(11)180926(21)206F7900; 206F7901 (01)04562122958974(11)180926(21)206F7901; 206F7936 (01)04562122958974(11)181113(21)206F7936; 206F7937 (01)04562122958974(11)181113(21)206F7937; 206F7940 (01)04562122958974(11)181113(21)206F7940; 206F7941 (01)04562122958974(11)181113(21)206F7941; 206F7942 (01)04562122958974(11)181113(21)206F7942; 206F7944 (01)04562122958974(11)181113(21)206F7944; 206F7945 (01)04562122958974(11)181113(21)206F7945; 206F7946 (01)04562122958974(11)181113(21)206F7946; 206F7947 (01)04562122958974(11)181114(21)206F7947; G3002552 (01)04562122958974(11)20190115(21)G3002552; G3024545 (01)04562122958974(11)190607(21)G3024545; G3037885 (01)04562122958974(11)20190719(21)G3037885; G3037887 (01)04562122958974(11)20190719(21)G3037887; G3042405 (01)04562122958974(11)190829(21)G3042405; G3042823 (01)04562122958974(11)20190807(21)G3042823; G3042828 (01)04562122958974(11)20190808(21)G3042828; G3045094 (01)04562122958974(11)20190822(21)G3045094; G3052105 (01)04562122958974(11)20191008(21)G3052105 G3052118 (01)04562122958974(11)20191025(21)G3052118 G3055549 (01)04562122958974(11)20191118(21)G3055549 G3055550 (01)04562122958974(11)20191118(21)G3055550 G3055552 (01)04562122958974(11)20191118(21)G3055552 G3055553 (01)04562122958974(11)20191118(21)G3055553 G3055554 (01)04562122958974(11)20191118(21)G3055554 G3055555 (01)04562122958974(11)20191118(21)G3055555 G3060234 (01)04562122958974(11)20191206(21)G3060234 G3060254 (01)04562122958974(11)20191206(21)G3060254 G3060275 (01)04562122958974(11)20191216(21)G3060275 G3060278 (01)04562122958974(11)20191216(21)G3060278 G3060690 (01)04562122958974(11)200127(21)G3060690 G3071074 (01)04562122958974(11)20200227(21)G3071074 G3080785 (01)04562122958974(11)201105(21)G3080785 G3090450 (01)04562122958974(11)201225(21)G3090450 G3110884 (01)04562122958974(11)210401(21)G3110884

Root Cause

Software design

Action Taken

Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21. Letter states reason for recall, health risk and action to take: Customers/Users are advised to accept the corrected software and manual as soon as available. Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems. We will contact you to set up time to update the software and provide the revised manual. (7) Other Information: For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email aaron.pierce@fujifilm.com

More recalls by Fujifilm Healthcare Americas Corporation

View all recalls by this company →