Cardinal Health Poly-Reinforced Surgical Gowns XX-large SKU: 9070
- Company
- Cardinal Health
- Recall Initiated
- November 2, 2021
- Recall Number
- Z-0374-2022
- Quantity
- 29,893 units
- Firm Location
- Waukegan, IL
Reason for Recall
Potential for open packaging seals compromising the sterility of the surgical gowns
Distribution
Worldwide distribution - US Nationwide and the countries of Chile, Thailand, Canada, Israel, LATAM Dist.Qatar, LATAM/Panama, LATAM/Mexico, LATAM/Colombia, LATAM/Costa Rica.
Lot / Code Info
Lot Numbers: 21CKQ001 21CKQ002 21HKQ001. UDI: 50192253005941 (cs) 10192253005943 (ea)
Root Cause
Process control
Action Taken
Cardinal Health issued Urgent Medical Device Recall Letter dated November 2, 2021, stating reason for recall, health risk and action to take: 1) REVIEW your inventory for affected product. 2) SEGREGATE and QUARANTINE all on-hand product that is confirmed to be labeled per Attachment 1. 3) NOTIFY any customers to whom you may have distributed or forwarded affected product about this product action. 4) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to gmb-fieldcorrectiveaction@cardinalhealth.com, whether or not you have affected product. 5) CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All other Customers 888-444-5440 6) CUSTOMERS that did not receive product directly from Cardinal Health should return product through the location where they purchased it. Available Assistance: Please contact the Customer Service group for any questions or to arrange for credit and return of any product: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All other customers 888-444-5440 For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.