VIDAS Varicella Zoster IgG (VZG) antibody IVD, kit

Company
Biomerieux, Inc.
Recall Initiated
December 10, 2019
Terminated
February 16, 2021
Recall Number
Z-1107-2020
Quantity
1,218 US units total (726 US, 492 OUS)
Firm Location
Durham, NC

Reason for Recall

Two batches of VIDAS Varicella-Zoster IgG (VZG ref 30217) may yield an invalid calibration which may delay the patient's sample result or a valid calibration may lead to a false positive patient result.

Distribution

Distributed nationwide and 9 units to Biomerieux Canada.

Lot / Code Info

US Model 30217-01 Lot 1007410330 and 1007393380

Root Cause

Unknown/Undetermined by firm

Action Taken

Users are instructed via customer letter to stop using the VIDAS VZG batches 1007410330 and 1007393380 and destroy the remaining products.