Boston Scientific Advanix Pancreatic Stent, Straight Leading Barb, 10F x 18cm UPN: M00536730
This recall has been terminated (originally issued November 25, 2019).
- Company
- Boston Scientific Corporation
- Recall Initiated
- October 21, 2019
- Posted
- November 25, 2019
- Terminated
- December 20, 2023
- Recall Number
- Z-0560-2020
- Quantity
- 35 units
- Firm Location
- Marlborough, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
Distribution
Nationwide Foreign: Austria Australia Belgium Bulgaria Croatia Czech Rep Denmark Finland France Germany Great Britain Italy Japan Pakistan Poland Russian Fed. Slovakia South Africa Spain Sweden Switzerland Thailand Turkey
Lot / Code Info
Lot Numbers: 21542436 21897084 23230412
Root Cause
Incorrect or no expiration date
Action Taken
Boston Scientific issued Notifications to consignees via Federal Express Priority mail on October 21, 2019. The notification instructs the account to check their inventory remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form.