RecallDepth

Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit number GUGE09I; e) LAP CHOLE PACK, kit number HNLC80I; f) Lap Chole Ovarian Appy, kit number JSLC06I; g) LAP CHOLE, ST LUKES, kit number LMLP08Z; h) ROBOTIC PACK - 242561, kit number MHRB88AJ; i) KIT, GENERAL LAPAROSCOPY, kit number MMLP40F; j) KIT, MAJOR LAPAROTOMY, kit number MMML26F; k) LAP CHOLE - LAP HERNIA PACK - 282807, kit number OWLC07K; l) GENERAL LAPAROSCOPY SAH, kit number SAGL57H; m) ROBOTIC PACK SAH, kit number SARB30K; n) ROBOTIC PACK SAH, kit number SARB30L; o) ROBOTIC PACK SAH, kit number SARB30M; p) LAP CHOLE PACK, kit number SMLC53; q) ROBOTICS PACK, kit number SNRP85D; r) ROBOTIC PACK, kit number SSRO22C; s) ROBOTIC PACK, kit number SSRO22D; t) TPK DAVINCI PROSTATE SYSTEM, kit number TPDV16; u) ASC LAP CHOLE PACK (PS 42279), kit number UICH53Y; v) GENERAL LAPAROSCOPIC PACK, kit number WVGL60E

This recall is currently active, issued January 26, 2024. It was issued by American Contract Systems, Inc..

Company
American Contract Systems, Inc.
Recall Initiated
November 29, 2023
Posted
January 26, 2024
Recall Number
Z-0822-2024
Quantity
3196 units
Firm Location
Temple Terrace, FL
Official Source
View on FDA website ↗

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Distribution

US Nationwide distribution.

Lot / Code Info

a) AGKD50L, UDI/DI 00191072141156, Lot Numbers: 905231, exp. 4/5/2024; 638221, exp. 12/28/2023; 689221, exp. 5/21/2023; b) ANPR34S, UDI/DI 00191072171351, Lot Numbers: 721221, exp. 10/6/2023; 879231, exp. 5/1/2024; 896231, exp. 4/14/2024; 927231, exp. 3/14/2024; 956231, exp. 2/13/2024; 645221, exp. 12/21/2023; c) BFLC02AD, UDI/DI 00191072169075, Lot Numbers: 654221, exp. 12/12/2023; 647221, exp. 12/19/2023; 987231, exp. 1/13/2024; UDI/DI 00191072188830, Lot Numbers: 946231, exp. 2/23/2024; 896231, exp. 4/14/2024; d) GUGE09I, UDI/DI none, Lot Numbers: 768221, exp. 8/20/2023; e) HNLC80I, UDI/DI 00191072146267, Lot Numbers: 797221, exp. 7/22/2023; 679221, exp. 11/17/2023; 995231, exp. 1/25/2024; 911231, exp. 3/30/2024; f) JSLC06I, UDI/DI 00191072126061, Lot Numbers: 839221, exp. 6/10/2023; 766221, exp. 8/22/2023; g) LMLP08Z, UDI/DI none, Lot Numbers: 661221, exp. 12/5/2023; 987231, exp. 1/13/2024; 942231, exp. 2/27/2024; 893231, exp. 4/17/2024; h) MHRB88AJ, UDI/DI none, Lot Numbers: 735221, exp. 9/22/2023; 710221, exp. 10/17/2023; 997231, exp. 1/3/2024; 959231, exp. 2/10/2024; 904231, exp. 4/6/2024; i) MMLP40F, UDI/DI none, Lot Numbers: 933231, exp. 3/8/2024; 879231, exp. 5/1/2024; j) MMML26F, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 896231, exp. 4/14/2024; k) OWLC07K, UDI/DI none, Lot Numbers: 682221, exp. 11/14/2023; 646221, exp. 12/20/2023; 982231, exp. 1/18/2024; 928231, exp. 3/13/2024; l) SAGL57H, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 790221, exp. 7/29/2023; m) SARB30K, UDI/DI none, Lot Numbers: 845221, exp. 6/4/2023; 888221, exp. 4/22/2023; n) SARB30L, UDI/DI none, Lot Numbers: 752221, exp. 9/5/2023; o) SARB30M, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 700221, exp. 10/27/2023; 792221, exp. 7/27/2023; 949231, exp. 2/20/2024; 983231, exp. 1/17/2024; p) SMLC53, UDI/DI 00191072188465, Lot Numbers: 969231, exp. 1/31/2024; q) SNRP85D, UDI/DI none, Lot Numbers: 666221, exp. 11/30/2023; 720221, exp. 10/7/2023; 793221, exp. 7/26/2023; 853221, exp. 5/27/2023; 965231, exp. 2/4/2024; r) SSRO22C, UDI/DI none, Lot Numbers: 891221, exp. 04/19/2023; s) SSRO22D, UDI/DI none, Lot Numbers: 660221, exp. 5/31/2023; 698221, exp. 10/29/2023; 756221, exp. 9/1/2023; 804221, exp. 7/15/2023; 840221, exp. 6/9/2023; 933231, exp. 8/31/2023; 987231, exp. 1/13/2024; t) TPDV16, UDI/DI 00191072128904, Lot Numbers: 856221, exp. 5/24/2023; 744221, exp. 9/13/2023; 675221, exp. 11/21/2023; 932231, exp. 3/9/2024; u) UICH53Y, UDI/DI none, Lot Numbers: 897231, exp. 4/13/2024; 882231, exp. 4/28/2024; v) WVGL60E, UDI/DI 00191072144034, Lot Numbers: 766221, exp. 8/22/2023; 694221, exp. 11/2/2023; 952231, exp. 2/17/2024;

Root Cause

Process control

Action Taken

Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.