Pediatric Heart SJH, REF SJPH34K

This recall is currently active, issued October 26, 2023. It was issued by American Contract Systems, Inc..

Company
American Contract Systems, Inc.
Recall Initiated
September 1, 2023
Posted
October 26, 2023
Recall Number
Z-0176-2024
Quantity
72 cases
Firm Location
Temple Terrace, FL
Official Source
View on FDA website ↗

Reason for Recall

During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.

Distribution

US Nationwide distribution in the states of FL, IA.

Lot / Code Info

UDI/DI 00191072167064, Lot Numbers: 975231

Root Cause

Process control

Action Taken

Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.