Revize; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x10 cm, 4x12 cm, 4x16 cm.
This recall is currently active, issued September 30, 2024. It was issued by Tei Biosciences, Inc..
- Company
- Tei Biosciences, Inc.
- Recall Initiated
- May 23, 2023
- Posted
- September 30, 2024
- Recall Number
- Z-2083-2023
- Quantity
- 199 units
- Firm Location
- Boston, MA
- Official Source
- View on FDA website ↗
Reason for Recall
Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing.
Distribution
Domestic: Nationwide Distribution.
Lot / Code Info
Item No. (UDI-DI): 611-001-001 (10381780357292), 611-001-002 (10381780357278), 611-001-003 (10381780357285).
Root Cause
Under Investigation by firm
Action Taken
The firm sent recall notifications to customers via email on 5/23/2023 and by mail on 5/24/2023. The recall notification informs customers that Integra is conducting a removal of all Surgimend PRS, Surgimend PRS Meshed, Surgimend, Surgimend MP, Revize, Revize-X, PriMatrix Dermal Repair Scaffold, and PriMatrix AG Antimicrobial Dermal Repair Scaffold devices due to identified issues with in-process and finished goods endotoxin testing that may result in out of specification endotoxin results. Customers are to check their stock for any lot of products listed in the provided Appendix and immediately remove them from service. All customers are asked to complete the provided acknowledgement form, regardless if affected product is on hand, and return it by email to Integralife6562@sedgwick.com or by fax to 888-229-0320. Questions about the acknowledgement form can be directed to 888-656-6375. Customers are to keep a copy of the recall notice for their records. A prepaid shipping label has been provided to all consignees so that return of products may be facilitated. Returns are being accepted for credit, however if a refund is preferred the firm asks that thsi is indicated in the "special instructions" section of the acknowledgement form. If any affected devices have been implanted, customers are to monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol. Customers with any questions can reach out to their Integra Sales Representative or Customer Service at 1-800-654-2873.