TriMed Countersink 1.7mm, HCS, REF HSINK-1.7, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.

This recall is currently active, issued September 28, 2018. It was issued by Trimed Inc..

Company
Trimed Inc.
Recall Initiated
September 28, 2018
Recall Number
Z-0355-2020
Quantity
99 instruments
Firm Location
Santa Clarita, CA
Official Source
View on FDA website ↗

Reason for Recall

The epoxy on Apex manufactured epoxy-coated instruments can pop off.

Distribution

Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.

Lot / Code Info

Lot number 91401455

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.