RecallDepth

TriMed Countersink 1.7mm, HCS, REF HSINK-1.7, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.

Company
Trimed Inc.
Recall Initiated
September 28, 2018
Recall Number
Z-0355-2020
Quantity
99 instruments
Firm Location
Santa Clarita, CA

Reason for Recall

The epoxy on Apex manufactured epoxy-coated instruments can pop off.

Distribution

Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.

Lot / Code Info

Lot number 91401455

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.

More recalls by Trimed Inc.

View all recalls by this company →