OEC FlexiView 8800, Digital Mobile Imaging System, GE Medical Systems
- Company
- Ge Oec Medical Systems, Inc
- Recall Initiated
- February 21, 2017
- Posted
- February 27, 2017
- Terminated
- April 24, 2017
- Recall Number
- Z-1469-2017
- Quantity
- 1
- Firm Location
- Salt Lake City, UT
Reason for Recall
OEC FlexiView 8800 workstation power cable assembly issue which can result in a loss of imaging functionality due to intermittent power cord failures.
Distribution
One consignee in CA.
Lot / Code Info
OEC FlexiView 8800 systems that had NEMA 5-15 15 Amp power cords replaced between January 28, 2016 and November 23, 2016.
Root Cause
Process control
Action Taken
On February 21, 2017, an Urgent Medical Device Correction letter was sent to one consignee with an affected unit; a field service representative hand delivered the customer communication.