Advincula Delineator with Ultem Plastic Soft Cup size 3.0 cm; Part No. AD750SC-KE30 The Advincula Delineator Uterine Manipulator is a single-use disposable device designed specifically for Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), and/or Laparoscopic Supra-Cervical Hysterectomy (LSCH).

Company
Coopersurgical, Inc.
Recall Initiated
June 14, 2019
Terminated
May 19, 2020
Recall Number
Z-1978-2019
Quantity
213 units in total
Firm Location
Trumbull, CT

Reason for Recall

The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.

Distribution

US distribution to state of: AZ, CA, GA, IN, LA, MA, NC, NH, NJ, NY, OH, PA, VA, and WA..

Lot / Code Info

Lot 021-19

Root Cause

Under Investigation by firm

Action Taken

On June 14, 2019, the firm, CooperSurgical, notified customers of the recall via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were informed that the soft, flexible cup of the Advincula Delineator may crack when subjected to excessive compression during use. The failure mode could lead to serious health consequences such as lacerations. Customers are asked to do the following; - Quarantine affected products - Complete the enclosed Acknowledgement and Receipt form and return to Cooper Surgical via email: recall@coopersurgical.com or fax to 203.601.9870 ATTN: Product Surveillance, even if you do not have affected product. - Once the firm has received your acknowledgement form, it will arrange for product returns and replacements, if requested. If you have questions regarding this recall, you may contact the firm at 2036015200 ext. 3300.