CS100 Intra-Aortic Balloon Pump, English, 110V; 0998-00-3013-53 For cardiovascular use.

Company
Datascope Corp.
Recall Initiated
June 17, 2019
Posted
July 16, 2019
Recall Number
Z-1941-2019
Quantity
16,656
Firm Location
Mahwah, NJ

Reason for Recall

If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being offered.

Distribution

Worldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND , AZERBAIJAN , IRAQ , PORTUGAL , BAHRAIN , IRELAND , QATAR , BANGLADESH , ISRAEL , ROMANIA , BELARUS , ITALY , RUSSIA , BELGIUM , JAPAN , SAUDI ARABIA , BOLIVIA , JORDAN , SERBIA , BOSNIA AND HERZEGOVINA , KAZAKHSTAN , SINGAPORE , BRAZIL , KENYA , SLOVAKIA , BRUNEI , KOSOVO , SLOVENIA , BULGARIA , KUWAIT , SOUTH AFRICA , CAMBODIA , LATVIA , SOUTH KOREA , CANADA , LEBANON , SPAIN , CHILE , LIBYA , SRI LANKA , CHINA , LITHUANIA , SWEDEN , COLOMBIA , LUXEMBOURG , SWITZERLAND , COSTA RICA , MACAU , SYRIA , CROATIA , MACEDONIA , TAIWAN , CYPRUS , MALAYSIA , TANZANIA , CZECH REPUBLIC , MALDIVES , THAILAND , DENMARK , MALTA , TRINIDAD , DOMINICAN REP , MARTINIQUE , TUNISIA , ECUADOR , MAURITIUS , TURKEY , EGYPT , MEXICO , TURKMENISTAN , EL SALVADOR , MOROCCO , UGANDA , ESTONIA , MYANMAR , UKRAINE , FINLAND , NAMIBIA , UNITED ARAB EMIRATES , FRANCE , NEPAL , UNITED KINGDOM , GEORGIA , NETHERLANDS , URUGUAY , GERMANY , NEW ZEALAND , UZBEKISTAN , GHANA , NIGERIA , VENEZUELA , GREECE , NORTH KOREA , VIETNAM , GUATEMALA , NORWAY , YEMEN , GUYANA , OMAN , ZAMBIA , HONDURAS , PAKISTAN , HONG KONG , and PALESTINE.

Lot / Code Info

All serial numbers affected

Root Cause

Use error

Action Taken

On June 17, 2019, the firm notified its customers of the product issue via Urgent Medical Device Correction letters. Customers are to take the following immediate interim actions: " Ensure the IABP is plugged into an AC power outlet whenever possible during patient use to prevent the battery from depleting. " Ensure the IABP is plugged into an AC power outlet when the system is not in use. The batteries should be kept at a full charge even when the IABP is not in use. " When transporting patients within or between facilities, please refer to the IABP Operating Instructions Manual for recommendations on portable/battery operation. For example: " Prior to portable operation, the battery should be fully charged " For Cardiosave Rescue and Cardiosave Hybrid only: " Additional charged batteries should be on hand during transport " Ensure the batteries are properly seated in the battery compartment/charger and the IABP Console is completely seated/secured into the IABP Cart " For Cardiosave Hybrid, you can verify if the Console is completely seated in the IABP cart by the indicator on the display. Check battery run time and replace batteries as required, as recommended in each IABPs Operating Instructions Manual. A reduction in run time can occur over a battery's life for reasons such as age, storage temperature and discharge depth. Batteries should be replaced: " After reaching the maximum number of charge-discharge cycles " When the battery provides less than the minimum specified run time " If the battery is broken, cracked, leaking or damaged " When the labeled lifetime of the battery is reached NOTE: Batteries for the Cardiosave Hybrid and Cardiosave Rescue IABPs sold before June 2015 should be replaced immediately as the labeled lifetime for these batteries is 4 years. Replacement batteries can be ordered through your sales or service representative. To determine the date of manufacture for all Cardiosave batteries, refer to attached d

More recalls by Datascope Corp.

View all recalls by this company →