Terumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19RE

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
June 14, 2019
Terminated
June 11, 2020
Recall Number
Z-2101-2019
Quantity
16 units
Firm Location
Elkton, MD

Reason for Recall

Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component

Distribution

MI Foreign: Japan, Belgium

Lot / Code Info

Lot Number: WP17

Root Cause

Process control

Action Taken

Terumo issued Urgent Medical Device Notification letter dated June 13 2019 via FedEx mailing executed on 6/14/2019 advising of the problem, health risk and action to take: Terumo CVS advises users to remove all affected product from inventory and return unused product to Terumo CVS following the instructions provided in this notice. Questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET Terumo Recall Fax: 1.410.392.7183 Terumo Recall Email: TCVSCustomerResponseElkton@terumomedical.com