Kii Fios First Entry, 5 X 75 mm, RX only, REF CFF05 Qty: 6, Sterile R, Lot: 1330981, Exp: July 15, 2021, Applied Medical Resource Corp. Product Usage: Kii Fios first entry is indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments, and as a means of insufflating the peritoneum prior to laparoscopic procedures.

This recall has been terminated (originally issued January 28, 2019).

Company
Applied Medical Resources Corp 22872 Avenida Empresa Rancho Santa
Recall Initiated
January 28, 2019
Terminated
June 12, 2020
Recall Number
Z-1959-2019
Quantity
2,016 units
Firm Location
Margarita, CA
Official Source
View on FDA website ↗

Reason for Recall

The product may not have met sterility requirements . Use of a non-sterile device on a patient may expose the patient to infectious agents.

Distribution

Worldwide Distribution - US Nationwide Distribution: OUS: Japan, Australia, New Zealand, Spain, Italy, Great Britain, Germany, France, and Finland

Lot / Code Info

Lot: 1330981, Exp: July 15, 2021 UDI: 0 0607915 12312 3

Root Cause

Packaging

Action Taken

Each US customer was notified via, "Urgent: Medical Device Recall" letter sent via UPS on January 28, 2019. All international consignees are in the process of being notified. The firm request customers compete the following actions: - Check your inventory for recalled product. - Complete the attached Recall Notification Confirmation Form (Page 2) to acknowledge the recall and indicate if your facility is returning or has already used product from the lot listed above. - If no product is being returned, please indicate on the Recall Notification Confirmation Form - If you are a distributor, please notify any facilities to which you distributed the affected product. Please also complete Page 3 of the Recall Notification Confirmation Form. - Return the Recall Notification Confirmation Form to Applied Medical by emailing it to recall60790124@appliedmedical.com or faxing it to (949) 713-8205. - Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical.