Kii Fios First Entry, 5 X 75 mm, RX only, REF CFF05 Qty: 6, Sterile R, Lot: 1330981, Exp: July 15, 2021, Applied Medical Resource Corp. Product Usage: Kii Fios first entry is indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments, and as a means of insufflating the peritoneum prior to laparoscopic procedures.

Company
Applied Medical Resources Corp 22872 Avenida Empresa Rancho Santa
Recall Initiated
January 28, 2019
Terminated
June 12, 2020
Recall Number
Z-1959-2019
Quantity
2,016 units
Firm Location
Margarita, CA

Reason for Recall

The product may not have met sterility requirements . Use of a non-sterile device on a patient may expose the patient to infectious agents.

Distribution

Worldwide Distribution - US Nationwide Distribution: OUS: Japan, Australia, New Zealand, Spain, Italy, Great Britain, Germany, France, and Finland

Lot / Code Info

Lot: 1330981, Exp: July 15, 2021 UDI: 0 0607915 12312 3

Root Cause

Packaging

Action Taken

Each US customer was notified via, "Urgent: Medical Device Recall" letter sent via UPS on January 28, 2019. All international consignees are in the process of being notified. The firm request customers compete the following actions: - Check your inventory for recalled product. - Complete the attached Recall Notification Confirmation Form (Page 2) to acknowledge the recall and indicate if your facility is returning or has already used product from the lot listed above. - If no product is being returned, please indicate on the Recall Notification Confirmation Form - If you are a distributor, please notify any facilities to which you distributed the affected product. Please also complete Page 3 of the Recall Notification Confirmation Form. - Return the Recall Notification Confirmation Form to Applied Medical by emailing it to recall60790124@appliedmedical.com or faxing it to (949) 713-8205. - Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical.