Innova Self-Expanding Stent System, 200 mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery (PPA) with reference vessel diameters from 4.0 mm to 7.0 mm and lesion lengths up to 190 mm. a. H74939293052030 (Material Description Innova 5 x 200 x 130); b. H74939293052070 (Material Description Innova 5 x 200 x 75); c. H74939293062030 (Material Description Innova 6 x 200 x 130); d. H74939293062070 (Material Description Innova 6 x 200 x 75); e. H74939293072030 (Material Description Innova 7 x 200 x 130); f. H74939293072070 (Material Description Innova 7 x 200 x 75); g. H74939293082030 (Material Description Innova 8 x 200 x 130); h. H74939293082070 (Material Description Innova 8 x 200 x 75)
- Company
- Boston Scientific Corporation
- Recall Initiated
- November 6, 2017
- Posted
- February 22, 2018
- Recall Number
- Z-0685-2018
- Quantity
- 22,916 units total
- Firm Location
- Maple Grove, MN
Reason for Recall
Stent possibly unable to be fully released from the delivery system.
Distribution
Worldwide Distribution - US Nationwide Distributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV. Distributed internationally to Canada
Lot / Code Info
a. Innova 5 x 200 x 130 (Batch No. 18095733, 18137885, 18233108, 18379812, 18362996, 18420576, 18404859, 18457350, 18478823, 18488703, 18519986, 18696468, 18670333, 18667082, 18756979, 18855752, 18844729, 18872178, 18886680, 18866631, 18892405, 18900294, 18951278, 19039704, 19064055, 19079251, 19111601, 19173251, 19206070, 19238815, 19231426, 19351863, 19309202, 19382422, 19367647, 19426956, 19483869, 19490837, 19501322, 19540549, 19540559, 19569664, 19641827, 19750808, 19742326, 19830060, 19844547, 19873365, 19907333, 19970577, 20014162, 20080095, 20096498, 20112175, 20186872, 20223208, 20216043, 20435880, 20151952, 20444934, 20439618, 20314154, 20336209, 20394072, 20608020, 20608137, 20607734, 20657170, 20420612, 20337854, 20450027, 20616298, 20777472, 20470018, 20672513, 20832034, 20876897, 20871401, 20728970, 20776633, 20851808, 20975844, 21020806, 21124105, 21045072, 21057925, 21091417, 21115593); b. Innova 5 x 200 x 75 (Batch No. 18343796, 18115802, 18379813, 19013283, 21124110); c. Innova 6 x 200 x 130 (Batch No. 18152540, 18137923, 18096697, 18119044, 18119040, 18115810, 18230863, 18108301, 18230864, 18211280, 18211281, 18245462, 18230637, 18230636, 18410450, 18428643, 18473268, 18283383, 18343799, 18439907, 18431333, 18368130, 18523235, 18519980, 18661526, 18488709, 18389662, 18498563, 18515152, 18446693, 18733063, 18759192, 18759189, 18766678, 19540741, 18791863, 18832544, 18792872, 18855754, 18860236, 18823216, 18866636, 18844738, 18892194, 18900365, 18870948, 18872541, 18878446, 18900381, 18888435, 18988348, 18961203, 19064043, 18945808, 19072771, 19111551, 19079253, 19151922, 19116475, 19164618, 19206072, 19226126, 19141574, 19245284, 19240145, 19288826, 19288832, 19309205, 19309206, 19367648, 19377502, 19408962, 19426948, 19389007, 19490838, 19511466, 19501332, 19517646, 19545788, 19555598, 19560848, 19569663, 19589370, 19584757, 19641826, 19635253, 19664685, 19654162, 19670616, 19709446, 19728413, 19699320, 19782868, 19693686, 19682956, 19682955, 19830064, 19873362, 19907334, 19905446, 19941738, 19942503, 19918783, 20013959, 19970579, 20014164, 20023939, 20109861, 20188996, 20080098, 20080097, 20188052, 20187538, 20080096, 20109863, 20112179, 20320562, 20830984, 20216045, 20183756, 20158213, 20433364, 20434473, 20233669, 20286988, 20233670, 20303729, 20342627, 20233668, 20607288, 20574727, 20607402, 20455108, 20368051, 20432603, 20511193, 20394775, 20594724, 20635843, 20776979, 20478648, 20821884, 20754076, 20758252, 20821886, 20800469, 20886809, 20825353, 20885852, 20938124, 20913156, 20950528, 20975845, 20997424, 21004412, 21119055, 21117395, 21058225, 21098042, 21071017, 21122202, 21145025, 21145024, 21219047, 21219242, 21219052); d. Innova 6 x 200 x 75 (Batch No. 18389661, 18623249, 18870949, 18900366, 19161485, 19245285, 19918786, 20635844, 20862432, 21098043); e. Innova 7 x 200 x 130 (Batch No. 18152541, 18099223, 18130828, 18113523, 18230638, 18217571, 18211282, 18186723, 18240225, 18168041, 18291418, 18339045, 18368242, 18283384, 18339044, 18316737, 18394605, 18385509, 18420569, 18313150, 18488713, 18921110, 18961213, 19072776, 19134318, 19173258, 19161488, 19206078, 19228862, 19255680, 19288830, 19309211, 19377507, 19327253, 19408970, 19436702, 19483875, 19491821, 19501423, 19527344, 19555961, 19545789, 19624864, 19659433, 19693528, 19670619, 19709445, 19830069, 19811429, 19742336, 19918785, 19907336, 19851529, 19974524, 19974523, 20080094, 20001802, 20109860, 20001803, 20188364, 20109862, 20177945, 20181650, 20210393, 20244728, 20321275, 20320380, 20286989, 20434033, 20267930, 20405389, 20381126, 20316570, 20574731, 20607286, 20352590, 20442151, 20466869, 20622461, 20655052, 20495971, 20518171, 20628127, 20777268, 20777261, 20820730, 20731022, 20833824, 20758256, 20886953, 20819295, 20856506, 20910181, 20937554, 21004415, 21119514, 20997428, 21119515, 21045505, 21103020, 21058247, 21138682, 21217652); f. Innova 7 x 200 x 75 (Batch No. 18336507, 18598712, 18733075, 18802238, 18892409, 18999955, 19151414, 19245286, 19511477, 19569668, 19684652, 19811430, 20699739, 20729796, 20978808, 21122996); g. Innova 8 x 200 x 130 (Batch No. 18211294, 18313156, 18385119, 18519981, 18583365, 18611781, 18733076, 18754584, 18780335, 18844724, 18977671, 19006855, 19048092, 19072779, 19188060, 19245289, 19304348, 19351866, 19454525, 19501424, 19624788, 19664694, 19684647, 19839023, 19782933, 20014160, 20080099, 20112342, 20216047, 20286991, 20439603, 20386608, 20655679, 20655677, 20329707, 20410534, 20655798, 20357647, 20997431, 21123358, 21123667); h. Innova 8 x 200 x 75 (Batch No. 18115809, 18313157, 18877978, 19161490, 19624794, 19887625, 20410535, 20862411)
Root Cause
Nonconforming Material/Component
Action Taken
Boston Scientific sent an Urgent Medical Device letter dated November 6, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers\ were instructed to immediately discontinue use of the product and remove all affected products from inventory, complete and return the Verification Form, return affected products and provide the notification to any healthcare professionals affected.