Biomerieux Vitek 2 Test kit VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

Company
Biomerieux, Inc.
Recall Initiated
March 15, 2019
Terminated
August 17, 2021
Recall Number
Z-1137-2019
Quantity
10784 cartons
Firm Location
Durham, NC

Reason for Recall

Unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits.

Distribution

Internationally to: United Kingdom, Latvia, Poland, Ukraine, Germany, and Scotland.

Lot / Code Info

Ref: 413723 (AST-N254), 418674 (AST-N330) and 422024 (AST-N371 Card) with 8.01/9.01 SW  Unique Device Identifier: 03573026404550 (AST-N254) / 03573026474836 (AST-N330) / 03573026604363 (AST-N371)

Root Cause

Software change control

Action Taken

On 03/15/2019, the firm, BioMerieux sent notifications to all their consignees with the following instructions: "Please take the following actions at this time: Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. Apply the VITEK 2 Systems bioART rule criteria as previously mentioned (i.e., using the criteria listed below for ESBL / E. coli isolates). Refer to Appendix A for detailed instructions. o If the card is AST-NXXX (where XXX indicates the card type designation, 254 / 330 / 371) o The organism is Escherichia coli o And the phenotype is ESBL o And the drug cefuroxime is <=8 o And the drug ceftazidime is <=0.5 o Then stop for consultation and add a comment: There is no evidence that the isolate is an ESBL producer. The category results for cephalosporins should be interpreted as tested, before the AES expertise. Please store this letter with your bioMrieux instrument documentation. Complete the Acknowledgement Form and return it to your local bioMrieux representative. bioMrieux, Inc. is committed to providing our customers with the highest quality products, and we apologize for any inconvenience this may have caused in your laboratory. If you have any questions or concerns, please contact your local bioMrieux representative."