RecallDepth

Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head

Company
Microport Orthopedics Inc.
Recall Initiated
June 21, 2016
Posted
December 1, 2016
Terminated
July 1, 2017
Recall Number
Z-0686-2017
Quantity
22 units
Firm Location
Arlington, TN

Reason for Recall

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Distribution

Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.

Lot / Code Info

REF 36693646 Serial Numbers: 09230967, 09229086, 10230508  REF 36693848 Serial Numbers: 08228745, 10230509, 11235075  REF 36694050 Serial Numbers: 09229091, 09230501  REF 36694252 Serial Numbers: 09229131, 10230510  REF 36694454 Serial Numbers: LLALMB161A, 09230503  REF 36694656 Serial Numbers: 09230505, 11230847  REF 36694858 Serial Numbers: 09229167, 10230511  REF 36695060 Serial Numbers: 09229169, 09230506  REF 36695262 Serial Numbers: 09230507, 09229170  REF 36695464 Serial Numbers: 08228747, 10230512

Root Cause

Device Design

Action Taken

MicroPort sent an Urgent Field Safety Notice dated June 22, 2016, to all affected customers notifying them of the change in the instructions for use on July 7th, 2016 by email. The firm also notified US hospitals where the implantation surgeries occurred within the last twelve (12) months by letter on July 13th, 2016 delivered by FedEx Priority. Customers were asked to complete and return the enclosed form even if they do not have any affected product to return by fax to 1-901-451-6032 or PostMarket@ortho.microport.com. For questions regarding this recall call 901-867-4771.

More recalls by Microport Orthopedics Inc.

View all recalls by this company →