Clear Flashback Needle 7cm x 18G AMC/4; Clear Flashback Needle 2.5 cm x 21G AMC/4; Winged Arterial Needle 7cm x 18G AMC/4; Clear Flashback Needle 7 cm x 19G AMC/4 Usage: Argon Arterial Access Needles are designed to provide vascular access for the percutaneous introduction of a guidewire.

This recall has been terminated (originally issued May 11, 2017).

Company
Argon Medical Devices, Inc
Recall Initiated
May 11, 2017
Terminated
January 25, 2018
Recall Number
Z-2252-2017
Quantity
7140 pieces
Firm Location
Athens, TX
Official Source
View on FDA website ↗

Reason for Recall

Argon Medical has received a complaint from one of their distributors of a potential packaging defect. The product is sealed in a pouch and the defect if present shows as a potential void or unsealed edge of the pouch.

Distribution

Worldwide Distribution - US Nationwide in the countries of Belgium, Canada, Japan, Netherlands

Lot / Code Info

11161099 11161729 11160990 11160912

Root Cause

Process design

Action Taken

Argon Medical Devices, Inc. is sending an Urgent Medical Device Recall letter by registered mail to the affected customer/hospital/distributor. Customers are instructed to inspect and quarantine and return any affected product remaining at their facility utilizing the instruction on the Recall Response Form.