Response BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000) is utilized to allow the surgeon to apply a recommended tension to the band and hold it in place during spine manipulation and set screw tightening.
- Company
- Orthopediatrics Corp
- Recall Initiated
- April 24, 2017
- Terminated
- September 18, 2017
- Recall Number
- Z-2321-2017
- Quantity
- 82
- Firm Location
- Warsaw, IN
Reason for Recall
Over-tensioning of the band resulting in damage to the band causing it to rupture
Distribution
Worldwide Distribution-US distribution to: MO, NY, KY, FL, NC and OH; and countries of: AUSTRALIA, FRANCE and SPAIN.
Lot / Code Info
All lots Product Number: 01-1102-1000
Root Cause
Device Design
Action Taken
The firm, OrthoPediatrics Corp., contacted each distributor/consignee via phone on 4/24/2017, regarding the recall. The phone call informed Consignees of issue and actions to be taken. The consignees were instructed to return of devices in their control were issued; cease further distribution of the product and quarantine it. The firm state that product will be distributed after Rework to new revision. If you have any questions, contact V.P. Regulatory Affairs, at (574) 268-6379 or email: mfox@orthopediatrics.com.