RecallDepth

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

Company
Carestream Health, Inc.
Recall Initiated
May 25, 2017
Posted
June 19, 2017
Terminated
June 13, 2018
Recall Number
Z-2651-2017
Quantity
Domestic: 1449 units; Foreign: 1416 units
Firm Location
Rochester, NY

Reason for Recall

Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.

Distribution

Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.

Lot / Code Info

Serial Numbers: Worldwide except China: 101 through 2623; China only: 800101 through 800482

Root Cause

Process control

Action Taken

Carestream Health sent the URGENT MEDICAL DEVICE CORRECTION Letters and Response Forms to the US consignees on May 25, 2017, via Certified Mail, Certified Receipt Requested. Letters to foreign accounts were distributed via e-mail or courier about the same time. Customers with questions were instructed to call 1-800-328-2910.

More recalls by Carestream Health, Inc.

View all recalls by this company →