Radiation Therapy Treatment Planning System, Model 5.0

This recall has been terminated (originally issued May 30, 2016).

Company
Raysearch Laboratories Ab Sveavaegen 9 Stockholm Sweden
Recall Initiated
May 30, 2016
Terminated
September 22, 2017
Recall Number
Z-0079-2017
Quantity
491 units (187 domestically & 304 internationally)
Official Source
View on FDA website ↗

Reason for Recall

For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.

Distribution

AZ, CA, FL, MT, NC NY, TX, OH & WA

Lot / Code Info

5.0.0.37, 5.0.1.11 and 5.0.2.35

Root Cause

Software design

Action Taken

On May 19th, 2016, RaySearch Laboratories distributed Field Safety Notice, Medical Device Correction notices and reply forms to their customers via email. Actions to be taken by the customer include: 1) Inspect all report templates that are intended to be used for plans with multiple beam sets. 2) Make sure that POI Dose statistics (Plan) is included immediately before ROI Dose statistics [Plan dose] in the Report content area. 3) Please educate planning staff and all users about this workaround. 4) Inspect your product and identify all installed units with the above software version number. Finally, complete and return the reply form via email to: freddie.cardel@raysearchlabs.com or by fax to: 888-501-7195 to confirm that you have read and understand the information. Customers with questions can contact David Hedfors via email: david.hedfors@raysearchlabs.com

More recalls by Raysearch Laboratories Ab Sveavaegen 9 Stockholm Sweden

View all recalls by this company →