The Spacelabs Healthcare Xhibit Telemetry Receiver (XTR), Model 96280 is used to transmit patient(s) physiological measurements and data to a monitor.

Company
Spacelabs Healthcare, Inc.
Recall Initiated
March 28, 2016
Terminated
December 19, 2016
Recall Number
Z-1502-2016
Quantity
217 units of XTRs and 531 units of PCBAs (164 XTRs and 425 PCBAs in US and 53 XTRs and 106 PCBAs outside the US)
Firm Location
Snoqualmie, WA

Reason for Recall

Reports of an abnormally high frequency of squelch events on Xhibit Telemetry Receivers (XTR), which is indicative of the loss of the telemetry data signal. In some cases, the Xhibit Central Monitor was also reported to display an Xhibit Offline error message for all patients being monitored.

Distribution

distributed in AL, AZ, CA, CO, CT, FL, ID, IN, KY, MI, MN, MO, MS, MT, SC, TX, WY and in the following countries: AUSTRALIA, CANADA, CHINA, FRANCE, NETHERLANDS, PHILIPPINES, ROMANIA, and UNITED KINGDOM.

Lot / Code Info

Affected PCBAs have no serial numbers. Below is the Part Numbers:  040-1736-00 KIT,UPGRADE,PCBA,QUAD RCVR,T BAND, 96280;  050-0758-00 REPLACEMENT KIT, PCBA,QUAD RCVR, Q BAND, 96280;  050-0759-00 REPLACEMENT KIT, PCBA,QUAD RCVR, T BAND, 96280; 050-0760-00 REPLACEMENT KIT, PCBA,QUAD RCVR, V BAND, 96280;  050-0758-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, Q BAND, 96280 EXCHANGE;   050-0759-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, T BAND, 96280 EXCHANGE;   050-0760-00-X REPLACEMENT KIT, PCBA,QUAD RCVR, V BAND, 96280 EXCHANGE;  050-0519-00 CM KIT,ET,600 MHZ-1400MHZ,NA, 96280; 050-0532-00 CM KIT,ET,400MHZ-600MHZ,INTERNATIONAL, 96280.  **************************SERIAL NUMBERS OF XTRs IN THE US************  6280-000051, 6280-000053, 6280-000056, 6280-000057, 6280-000058, 6280-000059, 6280-000066, 6280-000067, 6280-000068, 6280-000077, 6280-000078, 6280-000079, 6280-000080, 6280-000083, 6280-000084, 6280-000085, 6280-000086, 6280-000087, 6280-000088, 6280-000090, 6280-000092, 6280-000093, 6280-000095, 6280-000096, 6280-000097, 6280-000098, 6280-000099, 6280-000100, 6280-000101, 6280-000102, 6280-000103, 6280-000104, 6280-000105, 6280-000106, 6280-000107, 6280-000108, 6280-000109, 6280-000110, 6280-000111, 6280-000112, 6280-000113, 6280-000114, 6280-000115, 6280-000116, 6280-000117, 6280-000118, 6280-000119, 6280-000120, 6280-000121, 6280-000122, 6280-000123, 6280-000124, 6280-000125, 6280-000126, 6280-000127, 6280-000128, 6280-000129, 6280-000131, 6280-000132, 6280-000133, 6280-000135, 6280-000136, 6280-000137, 6280-000138, 6280-000140, 6280-000141, 6280-000142, 6280-000143, 6280-000145, 6280-000146, 6280-000147, 6280-000148, 6280-000149, 6280-000151, 6280-000152, 6280-000153, 6280-000154, 6280-000155, 6280-000156, 6280-000157, 6280-000158, 6280-000159, 6280-000160, 6280-000161, 6280-000167, 6280-000168, 6280-000169, 6280-000170, 6280-000171, 6280-000172, 6280-000173, 6280-000174, 6280-000175, 6280-000176, 6280-000177, 6280-000200, 6280-000227, 6280-000228, 6280-000229, 6280-000233, 6280-000234, 6280-000235, 6280-000236, 6280-000238, 6280-000239, 6280-000240, 6280-000241, 6280-000242, 6280-000243, 6280-000245, 6280-000248, 6280-000250, 6280-000251, 6280-000252, 6280-000253, 6280-000254, 6280-000255, 6280-000256, 6280-000257, 6280-000259, 6280-000260, 6280-000261, 6280-000263, 6280-000264, 6280-000265, 6280-000266, 6280-000267, 6280-000268, 6280-000269, 6280-000271, 6280-000272, 6280-000273, 6280-000274, 6280-000275, 6280-000278, 6280-000279, 6280-000280, 6280-000281, 6280-000283, 6280-000284, 6280-000285, 6280-000286, 6280-000287, 6280-000288, 6280-000289, 6280-000298, 6280-000299, 6280-000301, 6280-000302, 6280-000303, 6280-000304, 6280-000306, 6280-000307, 6280-000308, 6280-000309, 6280-000310, 6280-000311, 6280-000312, 6280-000313, 6280-000314, 6280-000315, 6280-000316, 6280-000319, and 6281-000144.     *********SERIAL NUMBERS OF XTRs OUTSIDE THE US************  6280-000048, 6280-000050, 6280-000052, 6280-000060, 6280-000061,  6280-000062, 6280-000063, 6280-000064, 6280-000065, 6280-000076,  6280-000094, 6280-000178, 6280-000179, 6280-000181, 6280-000182,  6280-000183, 6280-000184, 6280-000185, 6280-000186, 6280-000188, 6280-000189, 6280-000190, 6280-000191, 6280-000193, 6280-000194,  6280-000195, 6280-000196, 6280-000197, 6280-000198, 6280-000199,  6280-000202, 6280-000225, 6280-000226, 6280-000231, 6280-000232,  6280-000237, 6280-000246, 6280-000247, 6280-000249, 6280-000262,  6280-000270, 6280-000276, 6280-000282, 6280-000290, 6280-000291,  6280-000292, 6280-000293, 6280-000294, 6280-000295, 6280-000297,  6280-000300, 6280-000305, and 6280-000317.

Root Cause

Other

Action Taken

On March 28, 2016, a customer letter will be sent via priority service, return receipt requested, to all U.S. customers. On April 4, 2016, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Spacelabs recommends that you brief your staff regarding the failure modes listed above. If you continue to use affected Telemetry Receiver(s), please note: **When there is a loss of signal, a signal loss alarm is generated and a triangular-shaped squelch waveform is displayed for any channel exhibiting signal loss. If unacknowledged, this alarm will escalate to medium priority after five minutes. After an additional five minutes, this alarm will escalate to high priority. This failure is usually transient. **When the XTR goes offline, all telemetry waveform(s) associated with the XTR will disappear from the Central Station and an offline message will appear in each telemetry waveform zone. This failure is not recoverable. Please circulate this notification to all persons affected by the information it contains and add a copy to the appropriate monitor user manual until such time as corrections have been implemented by Spacelabs. Spacelabs will contact you to schedule a convenient time for Spacelabs to update your Xhibit Telemetry Receivers at no cost. If you have any questions about this corrective action, please contact Spacelabs at 1-800-522-7025 and select 2 for Technical Support.