RecallDepth

Patient Support Material Number: 9890 010 87431 used with: BuckyDiagnost ; Model: 704060/706037

This recall has been terminated (originally issued August 19, 2014).

Company
Philips Medical Systems, Inc.
Recall Initiated
August 19, 2014
Terminated
June 14, 2016
Recall Number
Z-0364-2016
Quantity
1517 systems, not all units have the patient support accessory
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.

Distribution

Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.

Lot / Code Info

All patient support for stitching are affected.

Root Cause

Device Design

Action Taken

The Field Safety Notice FCO70600086 will be mailed on August 19, 2014. Philips Medical Systems will send out certified letters and will have the receipts returned for tracking purposes as well as have a Field Service Engineer visit each site to provide the fix when the Field Change Order is released.

More recalls by Philips Medical Systems, Inc.

View all recalls by this company →