RecallDepth

Philips DigitalDiagnost Release 4.0.3, Release 4.1/4.1.1. excluding systems with SP1 Product Usage: The DigitalDiagnost is intended for use in generating radiographic images of human anatomy by qualified/trained doctors or technicians.

This recall has been terminated (originally issued March 23, 2015).

Company
Philips Electronics North America Corporation
Recall Initiated
March 23, 2015
Terminated
January 17, 2017
Recall Number
Z-2383-2015
Quantity
62
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.

Distribution

US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.

Lot / Code Info

DigitalDiagnost

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Philips Healthcare sent an Urgent Field Safety Notice letter dated June 25, 2015 to customers. The letter identified the affected product, problem and actions to be taken. Philips plans to install a software upgrade in affected systems, which will eliminate this issue. Philips Healthcare Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your local philips representative.

More recalls by Philips Electronics North America Corporation

View all recalls by this company →