Metasul Femoral Head - indicated for total hip arthroplasty for treatment of patient conditions of non-inflammatory degenerative joint disease.
This recall has been terminated (originally issued February 20, 2015).
- Recall Initiated
- December 17, 2014
- Posted
- February 20, 2015
- Terminated
- September 23, 2015
- Recall Number
- Z-1136-2015
- Quantity
- 16
- Official Source
- View on FDA website ↗
Reason for Recall
Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
Distribution
Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan
Lot / Code Info
Item numbers: 734028800 Lot #'s: 61487614
Root Cause
Package design/selection
Action Taken
On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.