Boston Scientific, Small Peripheral Cutting BalloonTM, Sterilized using irradiation: The Peripheral Cutting Balloon Catheters are indicated for Percutaneous Transluminal Angioplasty (PTA) of obstructive lesions of synthetic or native arteriovenous dialysis fistulae. 1) 4.00mm x 15mm Small Peripheral Cutting BalloonTM MonorailTM Microsurgical Dilatation Device, UPN #001BPM4015140F0, Catalogue # BPM4015140F & UPN # M001PCBM4015140F0, Catalogue # PCBM4015140F. 2) 4.00mm x 15mm Small Peripheral Cutting BalloonTM Over-the-Wire Microsurgical Dilatation Device, UPN # M001PCBO4015140F0, Catalogue # PCBO4015140F.
This recall has been terminated (originally issued September 3, 2013).
- Company
- Boston Scientific Corporation
- Recall Initiated
- July 31, 2013
- Posted
- September 3, 2013
- Terminated
- January 27, 2015
- Recall Number
- Z-2137-2013
- Quantity
- 1,331
- Firm Location
- Maple Grove, MN
- Official Source
- View on FDA website ↗
Reason for Recall
Boston Scientific is initiating a recall removal of the 4.0mm x 15mm size of Small Peripheral Cutting Balloon (Monorail and Over-the-Wire) Microsurgical Dilatation Devices and Flextome Cutting Balloon (Monorail and Over-the-Wire) Microsurgical Dilation Devices. Boston Scientific has determined that users may experience significant difficulty or inability to remove the protector cap from the devic
Distribution
Worldwide Distribution - USA including AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV,NH,NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC,SC, TN, TX, UT, VA, WA, WV, and WI. Internationally to Algeria, Finland, Japan, Spain, Australia, France, Latvia, Sweden, Austria, Germany, Netherlands, Switzerland, Belgium, Great Britain, Norway, Taiwan, Brazil, Greece, Panama, Turkey, Canada, Hong Kong, Poland, Unit.Arab Emir., China, India, Russian Fed., Colombia, Ireland, Saudi Arabia, Czech Republic, Israel, South Africa, Denmark, Italy, and South Korea.
Lot / Code Info
UPN # 001BPM4015140F0, Catalogue #BPM4015140F. Lot#'s 15778169, 15735527, 15774378, 15788826, 15812347, 15824636, 15825277, 15836626, 15873338, 15891705, 15931291, 15945240, 16057260. UPN# M001PCBM4015140F0, Catalogue #PCBM4015140F, Lot #15756145. UPN # M001PCBO4015140F0, Catalogue #PCBO4015140F, Lot #'s 15688199, 15992736, 16066536, 16141255.
Root Cause
Process change control
Action Taken
Boston Scientific sent an Urgent Medical Device Recall Removal letter dated July 30, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were advised that further distribution or use of the product should cease immediately, and to segregate the product and return it to Boston Scientific. Consignees were asked to complete and return the "Recall Removal Reply Verification Tracking Form". For questions consignees were instructed to contact their local Sales Representative. For questions regarding this recall call 763-494-2042.