Needle Counters with Blade Disarmer; 12.2 x 10.5 x 3.5 cm; 40 Count/70 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat. No. 4FA70SSA) Needle Counters with Blade Disarmer; 12.2 x 10.5 x 3.5 cm; 40 Count/70 Capacity; Sterile; Removable Numbered Foam Block; Magnetic Area (Cat. No. 4FM70SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

This recall has been terminated (originally issued March 28, 2013).

Company
Cardinal Health
Recall Initiated
December 18, 2012
Posted
March 28, 2013
Terminated
August 1, 2016
Recall Number
Z-0985-2013
Quantity
Cat. No. 4FA70SSA: 1,248; Cat. No. 4FM70SSA: 43,920
Firm Location
Mc Gaw Park, IL
Official Source
View on FDA website ↗

Reason for Recall

Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

Distribution

Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.

Lot / Code Info

Catalog Number: 4FA70SSA Lot Numbers: all lot numbers less than or equal to 931750  Catalog Number: 4FM70SSA Lot Numbers: all lot numbers less than or equal to 912173

Root Cause

Package design/selection

Action Taken

Cardinal Health sent an Urgent Product Recall letter dated December 18, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immedaitely stop using affected product and discard any product that may be on hand, return the enclosed acknowledgement form, notify any customers to whom you may have distributed product affected by this recall, and contact the appropriate Customer Service group to arrange for credit of any affected product discarded. For questions contact Cardinal Health Professional Services at 800-292-9332.