Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment.

This recall has been terminated (originally issued April 16, 2013).

Company
Accuray Incorporated
Recall Initiated
March 19, 2013
Posted
April 16, 2013
Terminated
October 23, 2013
Recall Number
Z-1126-2013
Quantity
176 units
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.

Distribution

Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.

Lot / Code Info

All units of these models.

Root Cause

Component design/selection

Action Taken

On 3/19/13, an "Urgent Device Correction Letter" was distributed to all affected users, with a description of the problem and user corrective action steps. US customers can contact 1-877- 668-8667 if questions.