RecallDepth

Nova StatStrip Xpress Glu-Test Strips (Glucose test strips) Catalog # 42214

Company
Nova Biomedical Corporation
Recall Initiated
February 25, 2013
Posted
April 19, 2013
Terminated
September 26, 2014
Recall Number
Z-1153-2013
Quantity
7433 cases
Firm Location
Waltham, MA

Reason for Recall

Glucose test strips reports no result message when tested

Distribution

Worldwide distribution: USA (nationwide) including states of: CA, IL, KY,NY, WV and countries of: United Kingdom and Italy.

Lot / Code Info

Lot 0312349249, Expires 12/31/2014 and Lot 0312361309, Expires 06/30/2015  Expanded recall: Lot 0313021309, Expires 07/31/2015

Root Cause

Process design

Action Taken

The firm, Nova Biomedical, sent an "Urgent Field Safety Notice 01-13SS" dated March 18, 2013 and US customers notification by phone contact and to the International customers including a Faxback Form to be communicated to and sent to Nova Biomedical Subsidiary by two methods of delivery (Email and Phone Call). The customers were instructed to INVALIDATE the Nova StatStrip Glucose Test Strip lots indicated in the list on all StatStrip meters within your institute; REPLACE all Nova StatStrip Glucose Test Strips from the indicated lots with replacement test strips provided by Nova; REMOVE all unused boxes of the Nova StatStrip Glucose Test Strip lots indicated from all locations within your institute; RETURN all unused boxes of the Nova StatStrip Glucose Test Strips lots indicated to Nova Biomedical U.K. or your local Nova distributor following the country specific instructions provided by your authorized Nova representative, and complete and return the Fax-Back Form via email or FAX to your Regional Service Manager at 781-8914806. On April 08, 2013, the firm expanded the recall notification to include one additional test strip lot #0313021309 which was shipped under two catalog numbers(Catalog # 42214 and Catalog# 51493). US Notification by phone contact with key customer contact facilitated by Senior Manger of Technical Support Services. International - Field Safety Notice and Faxback Form is being communicated to and sent to Nova Biomedical Subsidiary by two methods of delivery (Email and Phone Call). Questions regarding this notice should be directed to the following Nova Biomedical EU Authorized Representative at: Nova Biomedical Innovation House, Aston Lane South Runcorn,WA7 3FY Telephone: + 44 1244 287087 Email: Europe@novabiomedical.co.uk. or US at (781) 894-0800.

More recalls by Nova Biomedical Corporation

View all recalls by this company →