Accessory Cart for The ACHIEVA 3.0T and Panorama 1T Magnetic Resonance Imaging unit. Consists of diagnostic devices that produce cross sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body.

This recall has been terminated (originally issued December 21, 2010).

Company
Philips Medical Systems North America Co. Phillips
Recall Initiated
May 12, 2009
Posted
December 21, 2010
Terminated
December 28, 2010
Recall Number
Z-0794-2011
Quantity
3 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.

Distribution

Nationwide Distribution -- IL, MI, and MD.

Lot / Code Info

Site numbers: 505048, 505705, and 41445594.

Root Cause

Nonconforming Material/Component

Action Taken

On 05/12/2009, Philips began sending out the URGENT-Field Safety Notice MR Accessory Cart to their consignees. The letter identified the affected product and stated that the drawer rails were made from magnetic material and were not according specification. The hazards involved were also provided along with how to identify the affected product. The firm issued the FCO 78100296 and the firm's Field Service Engineer replaced the magnetic rails with the non-magnetic rails at the consignee's site. The consignees can contact contact the Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and reference FCO 78100296 with any communication.