Leksell Gamma Knife Perfexion. Radionuclide radiation therapy system. Elekta, Inc. Norcross, GA. Indicated for use in the stereotactic irradiation of intracranial structures.

This recall has been terminated (originally issued April 5, 2010).

Company
Elekta, Inc.
Recall Initiated
May 30, 2007
Posted
April 5, 2010
Terminated
December 28, 2010
Recall Number
Z-1231-2010
Quantity
18 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

There has been an issue with "Image Fushing" where low precision calculation caused images to become inaccurate.

Distribution

Nationwide Distribution -- AR, AZ, CA, FLA, IL, MI, MN, MO, MS, NJ, OH, OR, PA, UT and VA.

Lot / Code Info

Serial Numbers: 6004-6006, 6009-6012, 6014-6018, 6020-6023, 6025, 6027 and 6029-6032.

Root Cause

Software design

Action Taken

Elekta, Inc. issued a Field Change Order, Reference Number 0000946 dated March 14, 2008. Consignees were informed of the affected product and instructed on necessary steps to take. For further information, contact Elekta, Inc. at 1-770-300-9725.