Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.

This recall has been terminated (originally issued December 1, 2011).

Company
Elekta, Inc.
Recall Initiated
October 20, 2011
Posted
December 1, 2011
Terminated
August 9, 2012
Recall Number
Z-0342-2012
Quantity
123
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.

Distribution

Nationwide distribution: USA including states of: AL, AK, AZ, CA, CT, FL, GA, IL, IA, KY, MD, MS, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TX, VT, VA, WA, WV, WI, Puerto Rico, & DC.

Lot / Code Info

G30369, G29103, G34727, G32313, G30367, G29101, G31330, G34401, G32304, G30962, G30610, G30668, G29986, G28427, G28581, G28673, G28430, G28755, G28576, G28574, G28676, G28678, G28811, G28763, G28806, G28863, G28809, G28864, G28936, G29035, G29042, G30720, G28929, G29043, G29102, G28421, G29296, G29299, G29843, G31188, G29832, G29838, G29844, G29840, G29983, G30300, G30027, G29980, G30210, G30213, G30034, G30208, G30309, G30310, G30530, G30533, G30558, G30565, G32986, G30606, G30611, G30961, G30718, G30662, G30717, G31397, G31326, G31186, G31190, G31323, G29842, G34393, G31404, G31512, G31522, G31518, G31515, G31517, G30073, G32359, G32363, G32305, G32364, G32366, G32385, G32425, G32389, G32501, G32680, G32682, G32809, G32814, G34285, G32905, G32820, G32821, G32824, G32904, G32908, G33335, G33339, G33338, G33428, G33433, G34150, G34288, G34151, G34159, G34290, G34348, G34347, G34350, G34397, G34501, G34684, G34502, G34505, G34689, G30974, G34691, G34980, G34983, G34726

Root Cause

Nonconforming Material/Component

Action Taken

The firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.