Greatbatch OptiSeal Valved PTFE Peelable Introducer, Model 1000093-001, 1000093-002, 1000093-003, 1000093-004, 1000093-005, 1000093-006, 1000093-007. Sterilized using Ethylene Oxide. Greatbatch Medical, 2300 Berkshire Lane North, Minneapolis, MN 55441 USA. The OptiSeal Valved PTFE Peelable Introducer is designed to reduce blood loss and air intake by providing hemostatic sealing at venous pressures. The device is offered in two configurations, one without a sideport and one with a sideport attached to a segment of extension tubing terminating in a 3-way stopcock.

This recall has been terminated (originally issued August 27, 2010).

Company
Greatbatch Medical
Recall Initiated
July 2, 2010
Posted
August 27, 2010
Terminated
December 17, 2011
Recall Number
Z-2294-2010
Quantity
211 5-pack boxes
Firm Location
North Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

The outer sterile barrier (pouch) packaging for the 12 Fr device is not intact. No breaches in the inner sterile barrier (tray/tyvek lid) have been identified. The OptiSeal introducer and accessories sealed in the tray are sterile. Infection may result if a sterile field is contaminated by the outer surface of the inner tray.

Distribution

Worldwide Distribution: United States, including the states of MN and CA.

Lot / Code Info

Model / lot number: 1000093-001 / W1515452; 1000093-002 / W1545383; 1000093-003 / W1519579, W1524143, W1515454; 1000093-004 / W1524146, W1519578, W1520649, W1515441; 1000093-005 / W1515455; 1000093-006 / W1515456; 1000093-007 / W1560536.

Root Cause

Packaging

Action Taken

Greatbatch Medical issued an "Urgent Medical Device Recall letter dated July 12, 2010 via email to consignees. The letter described the problem and affected device. Consignees were instructed to immediately discontinue distribution of the product and return the unused units to Greatbatch Medical and to notify their customers to immediately return unused units. For shipping assistance, questions or assistance in notifying accounts, contact Shannon Springer at 763 951-8244 or Kimberly Briggs, 716 759-5702.