CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50

This recall has been terminated (originally issued September 29, 2014).

Company
Kci Usa, Inc.
Recall Initiated
May 19, 2014
Posted
September 29, 2014
Terminated
September 18, 2015
Recall Number
Z-2717-2014
Quantity
1,331 units
Firm Location
San Antonio, TX
Official Source
View on FDA website ↗

Reason for Recall

Blade can shift during shipping which makes the product ineffective.

Distribution

US nationwide Distribution

Lot / Code Info

Part number CT-H25 with the following lot numbers: 3244680000, 3446210000, 3246280000, 3246480000, 3246490000, 3246970000, 3247020000, 3247030000, 3247130000, 3247440000, 3247670000, 3247680000, 3247980000, 3248070000, 3248200000, 3248300000, 3249120000, 3249130000, 3249140000, 3249150000, 3249160000.  Part number CT-H50 with the following lot numbers: 324782, 324796, 324797, 3245120000, 3245670000, 3245680000, 3246270000, 3246490000, 3246690000, 3246700000, 3246940000, 3246980000, 3247140000, 3247160000, 3247170000, 3247250000, 3247310000, 3247470000, 324780000, 3247810000, 3247820000, 3247960000, 3247970000, 3248080000, 3248190000, 3248230000, 3248300000, 3248380000, 3248960000, 3248970000, 3248980000, 3248990000, 3249000000, 3249010000, 3249020000, 3249030000, 3249040000, 3249050000, 3249060000, 3249070000, 3249080000, 3249090000, 3249100000, 3249110000, 3249390000, 3249580000.

Root Cause

Device Design

Action Taken

KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions. Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s). Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4.