Fast-Cath Transseptal Guiding Introducer, 90, 8F, REF 406586, Length 60 cm, Sterile EO, Rx only, St. Jude Medical, 14901 DeVeau Place, Minnetonka, MN 55343 Indicated for introducing various cardiovascular catheters into the left side of the heart through the interatrial septum.

This recall has been terminated (originally issued December 28, 2012).

Company
St Jude Medical Inc
Recall Initiated
March 28, 2012
Posted
December 28, 2012
Terminated
July 12, 2013
Recall Number
Z-0617-2013
Quantity
26
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

St. Jude Medical is recalling a single batch of Fast Cath Transseptal Guiding Introducers which was packaged as a 180 degree curve style introducer instead of a 90 degree curve style introducer as indicated on the labeling.

Distribution

US Distribution only including the states of FL, KY and OK.

Lot / Code Info

Model no. 406586 Batch no. 3619533

Root Cause

Process control

Action Taken

St Jude Medical hand delivered the" Product Recall" letter dated March 27, 2012, to the affected customers on April 5, 2012, and March 28, 2012. The letter identified the product, the problem and the action needed to be taken by the customer. The returned devices and those not distributed were quarantined at the SJM Minnetonka distribution center until their destruction. Customers were requested to to complete and return the reconciliation form. Further questions please call (651) 756-6526