RecallDepth

Clinitron At Home Bed headboard and foot side rails. The Clinitron At Home Air Fluidized Therapy Unit provides a pressure-relieving sleep surface for patients in the home environment. It helps enable patients with pressure sores to heal their wounds

This recall has been terminated (originally issued March 1, 2012).

Company
Hill-Rom, Inc.
Recall Initiated
February 3, 2012
Posted
March 1, 2012
Terminated
December 5, 2012
Recall Number
Z-1131-2012
Quantity
Foot Siderail = 70; Headboards = 103
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

The firm identified in October 2011 that the Clinitron Rite Hite Bed and Clinitron At Home Bed headboards failed to meet the safety standard force requirements. The firm identified in October 2011 that the Clinitron Rite Hite Bed and Clinitron At Home Bed outer rail, which is extruded Aluminum tubing, is intended to be one piece joined together by a weld at the lower center. The parts receiv

Distribution

Worldwide Distribution -- USA (nationwide) including the states of AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY. and the country of Canada.

Lot / Code Info

The headboard (p/n 6020200003) and siderail (p/n 6020200002) are not serialized. There are no model, catalog or code number identification marks available on the accessories. It is possible to identify these accessories though by a mark stamped into the mounting post that are used to attach these accessories to the bed. May 2011 and October 2011 headboards and siderails were shipped to customers.

Root Cause

Nonconforming Material/Component

Action Taken

Hill-Rom, Inc. sent a Modification Notice Urgent Medical Device Correction Notification dated February 2, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The modification notice instructed all customers to find and replace all headboards and food side rails that do not meet design intent. All technicians will inspect all affected product in the service centers and all at home and rite hite beds for an "x" or "L" stamp. The technicians will inspect the beds that are still in use during scheduled maintenance or before each rental. All will be completed before 01-JUL-12. For support call 1-800-445-3720

More recalls by Hill-Rom, Inc.

View all recalls by this company →