HAMILTON G5 ventilators with software version V2.0x. Hamilton Medical AG, Switzerland Imported and distributed by Hamilton Medical, 4990 Energy Way, Reno, NV 89502. Product Usage: Hamilton-G5 Ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The devices are intended for use in the hospital and institutional environment with trained healthcare professionals.

Company
Hamilton Medical Inc
Recall Initiated
February 6, 2012
Posted
March 7, 2012
Terminated
October 23, 2013
Recall Number
Z-1185-2012
Quantity
456 devices in US.
Firm Location
Reno, NV

Reason for Recall

Hamilton-G5 ventilators with software version V2.0x may experience an unintentional change of the displayed ventilation mode under certain conditions.

Distribution

(USA) Nationwide Distribution including the states of: CA, CO, FL, GA, MA, MI, MO, NC, NV, NY, PA, TN, TX, VA and WV.

Lot / Code Info

All Hamilton 5

Root Cause

Software design

Action Taken

Hamilton Medical sent a Medical Device Safety Alert and Correction Action letter dated February 17, 2011 to all affected customers via a traceable method. The letter identified the affected product, problem and immediate actions required by operators, distributor and manufacturer. The letter states that a Hamilton Medical Account manager will contact customers to schedule a convenient time for the Technical Support Department to install the software update. Questions or concerns may be directed to Technical Support Manager at 800-426-6331, Ext 125.