RecallDepth

BladderScan BVM 9500 The BladderScan BVM 9500 is intended to project ultrasound energy through the lower abdomen to obtain an image of the bladder and measure urinary bladder volume and mass noninvasively.

This recall has been terminated (originally issued October 4, 2010).

Company
Verathon, Inc.
Recall Initiated
August 23, 2010
Posted
October 4, 2010
Terminated
September 8, 2011
Recall Number
Z-0005-2011
Quantity
260 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

The BladderScan BVM 9500 devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.

Distribution

Worldwide Distribution - USA and the countries of Australia, Canada, China, Colombia, European Union, United Kingdom, Saudi Arabia, and Singapore.

Lot / Code Info

Serial numbers:   B5001000, B5001003, B5001004, B5001007, B5001011, B5001016, B5001017, B5001018, B5001020, B5001021, B5001022, B5001023, B5001024, B5001025, B5001026, B5001027, B5001032, B5001062, B5001069, B5001070, B5001074, B5001075, B5001076, B5001077, B5001078, B5001079, B5001080, B5001081, B5001088, B5001090, B5001091, B5001092, B5001093, B5001095, B5001098, B5001099, B5001101, B5001102, B5001103, B5001104, B5001105, B5001106, B5001107, B5001108, B5001109, B5001110, B5001111, B5001113, B5001114, B5001115, B5001116, B5001117, B5001118, B5001119, B5001120, B5001121, B5001122, B5001123, B5001124, B5001125, B5001126, B5001127, B5001128, B5001129, B5001132, B5001138, B5001139, B5001140, B5001141, B5001142, B5001143, B5001144, B5001147, B5001148, B5001149, B5001150, B5001152, B5001156, B5001157, B5001158, B5001159, B5001160, B5001162, B5001164, B5001166, B5001170, B5001171, B5001172, B5001173, B5001174, B5001175, B5001176, B5001177, B5001179, B5001180, B5001181, B5001182, B5001184, B5001186, B5001187, B5001188, B5001189, B5001190, B5001191, B5001192, B5001194, B5001195, B5001197, B5001198, B5001199, B5001200, B5001201, B5001202, B5001203, B5001204, B5001206, B5001208, B5001209, B5001210, B5001211, B5001212, B5001215, B5001217, B5001218, B5001219, B5001220, B5001221, B5001222, B5001223, B5001224, B5001225, B5001226, B5001228, B5001230, B5001231, B5001233, B5001234, B5001235, B5001236, B5001237, B5001238, B5001239, B5001240, B5001241, B5001245, B5001246, B5001247, B5001249, B5001250, B5001251, B5001252, B5001253, B5001254, B5001256, B5001264, B5001265, B5001266, B5001267, B5001268, B5001269, B5001271, B5001273, B5001274, B5001275, B5001276, B5001277, B5001278, B5001279, B5001280, B5001282, B5001283, B5001285, B5001286, B5001288, B5001289, B5001290, B5001291, B5001292, B5001293, B5001294, B5001295, B5001296, B5001297, B5001298, B5001299, B5001300, B5001302, B5001309, B5001310, B5001313, B5001314, B5001315, B5001316, B5001317, B5001318, B5001319, B5001320, B5001321, B5001322, B5001323, B5001327, B5001328, B5001329, B5001330, B5001331, B5001332, B5001333, B5001334, B5001335, B5001336, B5001338, B5001339, B5001340, B5001341, B5001342, B5001343, B5001344, B5001345, B5001346, B5001347, B5001348, B5001349, B5001350, B5001351, B5001352, B5001353, B5001354, B5001355, B5001356, B5001357, B5001358, B5001359, B5001360, B5001363, B5001364, B5001365, B5001366, B5001367, B5001368, B5001369, B5001370, B5001371, B5001372, B5001373, B5001374, B5001375, B5001376, B5001377, B5001378, B5001380, B5001383, B5001384, B5001385, B5001386, B5001387, B5001388, B5001390, B5001391, B5001392, and B5001393.

Root Cause

Software design

Action Taken

Verathon sent an URGENT Medical Device Recall Notification letter dated August 12, 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Verathon provided an addendum to the User Manual with instructions on how to reset the instrument. Customers were advised to: " Add the enclosed addendum to their current Users Manual. " Update the application software of the device to v2.7.0 or higher. If the customers had questions or needed support for the software update process, they were to contact Verathon Medical Customer Care at (425) 867-1348 or (800) 331-2313 or cservice@verathon.com.

More recalls by Verathon, Inc.

View all recalls by this company →