Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.

This recall has been terminated (originally issued September 1, 2010).

Company
Elekta, Inc.
Recall Initiated
September 1, 2010
Terminated
December 14, 2010
Recall Number
Z-0350-2011
Quantity
23 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

It is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode.

Distribution

Nationwide distribution: AZ, CA, DC, IL, KY, MA, MI, MS, NV, NY, OH, OR, PA, VT, and WV.

Lot / Code Info

Serial Numbers 152019, 151866, 151259, 151522, 151969, 152192, 152126, 152115, 152116, 152176, 151931, 152177, 151523, 151626, 151842, 152158, 152079, 152080, 151250, 152064, 152207, 151628, and 151362

Root Cause

Other

Action Taken

Elekta sent an "Important Notice A340" dated September 1, 2010 to all customers. The notice described the product, problem and actions to be taken by the customer. The customers were instructed to make a decision if XVI or MOSAIQ does the corrective table movement; configure and ensure that the system only uses RATM or CMA for the corrective table movement, when in SYNERGISTIQ mode, and complete and return the IMPORTANT NOTICE Confirmation of Receipt form via fax at +44(0)1293 654401. If you have any questions contact your local service representative at +44(0)1293 654200.