Leksell Gamma Knife C

This recall has been terminated (originally issued April 29, 2010).

Company
Elekta, Inc.
Recall Initiated
September 1, 2006
Posted
April 29, 2010
Terminated
December 14, 2010
Recall Number
Z-1533-2010
Quantity
88 units
Firm Location
Norcross, GA
Official Source
View on FDA website ↗

Reason for Recall

Leksell Gamma Knife C 1.2 with MCU version 4.0.0.6 improperly adjusted trolley for 4mm helmet caused jamming when helmet changer attempted to lock helmet. Lock did not fully engage and could still be raised. When the lock does not engage, then when the helmet changer rises, the helmet could disengage from the helmet changer and fall back onto the helmet trolley.

Distribution

Nationwide, Austria, Belgium, Canada, China, Croatia, Italy, Japan, Korea, Norway, Spain, Sweden, Taiwan and Turkey.

Lot / Code Info

4149, 4150, 4151, 4157, 4158, 4160, 4151, 4165, 4168, 4171, 4175, 4176, 4178, 4183, 4187, 4189, 4191, 4193, 4194, 4195, 4196, 4197, 4199, 4201, 4202, 4204, 4207, 4208, 4212, 4300, 4302, 4303, 4306, 4307, 4308, 4309, 4313, 4314, 4315, 4318, 4319, 4321, 4322, 4326, 4328, 4329, 4334, 4337, 4802, 4809, 4816, 4817, 4818, 4824, 4828, 4830, 5000, 5003, 5006, 5009, 5010, 5001, 4803, 4310, 4210, 4311, 4330, 5004, 5014, 4305, 4320, 4317, 4324, 4325, 5011, 5012, 4304, 4316, 5002, 5007, 5008, 5013, 4819, 4806, 5005, 4301, 4323, 4312

Root Cause

Other

Action Taken

Phase 1 - Field Change Order FCO-DOC-06-0003 "Technical Note - Potential risk of dropping the collimator helmet" (dated Aug. 28, 2006) was sent to affected customers as notification of the problem. Phase 2 - Technical Notes Leksell Gamma Knife C version 1.2 (dated July 2006) "Important Notice - Potential risk of dropping the collimator helmet" was sent to all Elekta Service Engineers as notification of the problem. The firm planned to create an upgrade kit for the effected Leksell Gamma Knife C1.2 systems. A service representative from Elekta will contact each site to decide on a suitable implementation schedule.