Revision K Actuator Test Boards sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 190713 Revision K

This recall has been terminated (originally issued May 13, 2014).

Company
Fresenius Medical Care Holdings, Inc.
Recall Initiated
April 25, 2014
Posted
May 13, 2014
Terminated
May 18, 2021
Recall Number
Z-1619-2014
Quantity
39 units
Firm Location
Waltham, MA
Official Source
View on FDA website ↗

Reason for Recall

2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump

Distribution

USA (nationwide) and the country of Canada.*

Lot / Code Info

Revision K Actuator Test Boards

Root Cause

Device Design

Action Taken

Fresenius Medical issued Urgent Recall letter issued on 4/25/14 by certified mail with signature confirmation and fax-back form. Customers instructed to examine their inventory to determine whether they have any of the affected 2008T machines or if they have installed any of the affected replacement boards. If customers have the affected machines or replacement boards, they will be instructed to contact FMC-RTG Technical Service to have the Actuator Test Boards replaced. Contact the FMCNA Technical Service Team at 1-800-227-2572.