Bio-Rad Variant II Hemoglobin Testing System with CDM v 4.0, model number 270-2000, Manufactured by Bio-Rad Laboratories, Hercules, CA Measurement of percent hemoglobin Alc is effective in monitoring long-term glucose control in individuals with diabetes mellitus.

Company
Bio-Rad Laboratories Inc
Recall Initiated
December 9, 2008
Posted
May 19, 2009
Terminated
November 12, 2010
Recall Number
Z-1395-2009
Quantity
42 units
Firm Location
Hercules, CA

Reason for Recall

The wrong barcode number and tube position number may be assigned to samples leading to discrepant sample results.

Distribution

Worldwide Distribution -- USA, Argentina, Australia, Brazil, Canada, China, Czech Republic, France, Germany, Hungary, India, Italy, Mexico, New Zealand, Singapore, South Korea, Sweden, Thailand, and the UK.

Lot / Code Info

CDM version 4.0 software

Root Cause

Pending

Action Taken

Recall initiated on December 10, 2008. Notification conducted by phone call and fax for domestic consignees. Each subsidiary will receive the subsidiary Urgent: Medical Device Correction notice, customer medical device correction response form and number of customers distributed to by e-mail. Each subsidiary must contact customers and provide the customer notice and response form as needed. The notice stated that Bio-Rad Laboratories would release a follow up communication providing a permanent solution upon completion of an investigation. In the interim, it was recommended that customers carefully review the Summary Report for duplicate barcode IDs or or tube position numbers within the same rack. If there are any questions or assistance needed, the local regional Bio-Rad office should be contacted.