Gravity Compensating Accessory; High Pressure Range, Sterile, Single Use, Rx only. CSF shunt accessory. Manufacturer: Integra Neuro Sciences Implants, SA, Cedex, France; distributed by: Integra NeuroSciences, Plainsboro, NJ 08536 The GCA is an implantable device designed to be used in conjunction with implanted systems which shunt cerebrospinal fluid CSF from the cerebral ventricles to an appropriate drainage site.

This recall has been terminated (originally issued December 12, 2008).

Company
Integra Lifesciences Corp.
Recall Initiated
October 3, 2008
Posted
December 12, 2008
Terminated
October 6, 2009
Recall Number
Z-0458-2009
Quantity
49 units
Firm Location
Plainsboro, NJ
Official Source
View on FDA website ↗

Reason for Recall

Integra NeuroSciences has become aware that certain Gravity Compensating Accessory implants have the potential for CSF leakage under certain conditions.

Distribution

Worldwide Distribution - MI, IL, NY, IN, CA, IN and NC and Carnegie, Victoria, Melbourne, Australia.

Lot / Code Info

Catalog number: 903-430, Lot Numbers 148792, 149247; Catalog Number 903-435, Lot Numbers 147796, 148790; Catalog Number 903-440; and Lot Numbers 147795, 149246

Root Cause

Process control

Action Taken

Urgent: Product Recall Notification letters, dated 10/3/08, were sent out on October 9, 2008. Letters explain the reason for the recall and the catalog and lot numbers affected. If the product has been implanted, it is recommended that the patient be monitored periodically using the standard of care determined appropriate by the patient's attending physician. If the product is found in inventory, customers are asked to quarantine the product and contact their Integra Sales Rep for assistance and for arrangements for product return. A Recall Acknowledgement and Return Form is to be completed and returned to Integra for reconciliation purposes.