Sterrad CycleSure Biological Indicator, Product Code 14324. The Sterrad CycleSure Biological Indicator is intended to be used as a standard method for frequent monitoring of the Sterrad Sterilizer cycles.

This recall has been terminated (originally issued October 20, 2009).

Company
Advanced Sterilization Products
Recall Initiated
June 23, 2009
Posted
October 20, 2009
Terminated
November 29, 2011
Recall Number
Z-0066-2010
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Biological Indicator indicating a positive result does not maintain yellow color for the entire incubation period and some indicators experience media evaporation prior to the end of the labeled maximum incubation time.

Distribution

Worldwide Distribution.

Lot / Code Info

All Lots.

Root Cause

Device Design

Action Taken

An Important Product Labeling Correction - New Instructions for Use dated, June 26, 2009, has been proposed to be sent by the firm for Cyclosure Biological Indicators to consignees. The revised instructions reduce the maximum read time for the indicators to 3 days (72 hours). The letter also states why ASP changed the Instructions. Questions are to be directed to ASP Customer Care Center.