RecallDepth

ROC Lumbar bolt caddy, Part Number: 81997-01-05. Sterilization container for ROC bolts/screws. Product is used to contain spinal implants.

This recall has been terminated (originally issued October 8, 2009).

Company
Alphatec Spine, Inc.
Recall Initiated
December 16, 2008
Posted
October 8, 2009
Terminated
November 13, 2009
Recall Number
Z-0036-2010
Quantity
7 Screws
Firm Location
Carlsbad, CA
Official Source
View on FDA website ↗

Reason for Recall

Alphatec discovered that the lot of bolt caddies had the incorrect bolt length laser etched on one side of the caddy. The actual ROC bolts themselves are correctly laser etched.

Distribution

Nationwide Distribution -- Including states of TX, FL, & NJ.

Lot / Code Info

Lot Number: 53298

Root Cause

Employee error

Action Taken

Alphatec contacted consignees via phone on December 16, 2008 and informed consignees that they will arrange to have Federal Express contact them for pick-up. Consignees were asked to check their inventory to see where this product is located, and ensure that it is put on hold for return; and what days and hours they are in the office, (Confirm address: attach copy of the pick-up request form.) If they have any follow up questions, they were told to call 1-800-922-1356.