Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.0 x 43MM, spinal implant component

This recall has been terminated (originally issued November 24, 2009).

Company
Stryker Spine
Recall Initiated
October 2, 2009
Posted
November 24, 2009
Terminated
January 23, 2013
Recall Number
Z-0430-2010
Quantity
181
Firm Location
Allendale, NJ
Official Source
View on FDA website ↗

Reason for Recall

Surgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break at the tip.

Distribution

Nationwide

Lot / Code Info

Catalog number 48036043, all lots

Root Cause

Device Design

Action Taken

Recall notification letters were sent to all Stryker branches, Hospital Risk Management and Surgeons on October 19, 2009 by Federal Express. Questions are to be directed to Tiffani Rogers, Regulatory Compliance Manager at 201-760-8206.