LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit Catalog #3121 Simulfluor Flu A / Flu B DFA kit.

This recall has been terminated (originally issued March 6, 2009).

Company
Millipore Corporation
Recall Initiated
November 24, 2008
Posted
March 6, 2009
Terminated
August 10, 2012
Recall Number
Z-0954-2009
Quantity
258 units
Firm Location
Temecula, CA
Official Source
View on FDA website ↗

Reason for Recall

Certain lots of Simulfluor Flu A / Flu B Reagent may cross-react with Respiratory Synctial Virus. The continued use of Simulfluor Flu A / Flu B Reagent Catalog #5250 may lead to a potential misdiagnosis of influenza B.

Distribution

Worldwide Distribution -- USA including states of AL, CA, CO, CT, FL, ID, IL, KS, MA, MI, MS, NM, NY, OH, OK, OR, PA, SD, TX, UT, VA, VT, WA, WV, and WI and countries of Canada, Mexico, Italy, Belgium,Germany, France, Spain, and United Kingdom.

Lot / Code Info

Lot Numbers JBC1531414, TR1458084, TR1480746, JBC1531421, JH1534500, TR1501923, TR1417106, and TR1422506.

Root Cause

Pending

Action Taken

Millipore is performing a voluntary recall by notifying customers and distributors by telephone on November 24, 2008. In addition, an Urgent: Medical Device Recall notification letter with the same date was sent via certified mail. Customers were advised to examine their stock immediately and discontinue the use of the specified lots. Customers were instructed to complete the enclosed Effective Check Survey confirming disposal of defective lots and return this form to Millipore's Regulatory Department. Customer service is to be contacted at 1-800-437-7500 for replacement product. Millipore Corporation will issue credit or replace the recalled lots that customers currently have in inventory.