Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Conical Reamer, instrument for shoulder arthroplasty, REF no. 00-4309-021-00.
This recall has been terminated (originally issued March 29, 2007).
- Company
- Zimmer Inc.
- Recall Initiated
- February 1, 2007
- Posted
- March 29, 2007
- Terminated
- May 4, 2007
- Recall Number
- Z-0675-2007
- Quantity
- 4
- Firm Location
- Warsaw, IN
- Official Source
- View on FDA website ↗
Reason for Recall
The reamer may lock up on the threads of the reamer body rather than spin freely as intended by design.
Distribution
California, Ohio and Texas.
Lot / Code Info
Lot 60605829.
Root Cause
Other
Action Taken
Zimmer sales staff was notified by letter dated 2/1/07 to pull any of the product from stock and to provide the name and address of their hospital customer accounts for further notification by Zimmer.