Large-Volume IV Administration Sets for use with Deltec 3000 and Deltec Micro 3100 Large Volume Infusion Pumps. Model Number 8C290 (Reorder No. 21-0304-01) Speciality Administration Set. Sterile, non-pyrogenic. Length 105 inch/267 cm, 25 ml priming volume, 20 drops/ml and 0 Injection sites, PVC cassette, non-PVC lined tubing, non-PVC drip chamber. Manufactured for Deltec SIMS Deltec, Inc., St. Paul, MN 55112 U.S.A. sims Smiths Industries Medical Systems.
- Company
- Smiths Medical Md, Inc.
- Recall Initiated
- December 8, 2005
- Posted
- March 18, 2006
- Terminated
- January 5, 2007
- Recall Number
- Z-0660-06
- Quantity
- 7 boxes of 30 Admin. Sets (210 sets in total)
- Firm Location
- Saint Paul, MN
Reason for Recall
Inaccurate infusion rate-administration sets which are used with selected infusion pumps may have a defect which does not allow the pump valve to close completely.
Distribution
South Carolina, Australia, Canada and Puerto Rico.
Lot / Code Info
Lot number FEB04T06
Root Cause
Other
Action Taken
Affected customers were notified of the recall and asked to return or destroy any unused product.