VacLok Syringe. Label reads in part: ''VACLOC SYRINGE 60 ml VACLOF SYRINGE-FIXED MALE LUER ... MERIT MEDICAL''. Sterile if package is unopened or undamaged. Part #VAC160.

This recall has been terminated (originally issued November 19, 2003).

Company
Merit Medical Systems, Inc
Recall Initiated
October 1, 2003
Posted
November 19, 2003
Terminated
February 2, 2004
Recall Number
Z-0139-04
Quantity
287 units
Firm Location
Jordan, UT
Official Source
View on FDA website ↗

Reason for Recall

There is a potential for the PETG blister trays to bond together with varying degrees of attachment. The potential for non-sterility exists when the attached trays are seperated by force and holes or cracks are generated in the seperated trays.

Distribution

Distributed nationwide. International distribution includes Canada, Japan, Korea, England and France.

Lot / Code Info

Lot # A282003

Root Cause

Other

Action Taken

Merit notified sales representatives, domestic and international accounts and dealers immediately by letter dated October 3, 2003. The manufacturer requests that the use of the affected inventory should immediately cease and that on-hand units should be promptly isolated from general inventory for simultaneous inspection by a site representative and a Merit sales representative.